Artifact in central venous pressure waveform due to central venous catheter tip abutting the wall of superior vena cava |
|
GN Chennakeshavallu1, S Sruthi1, Saravana Babu2
1 Division of Cardiothoracic and Vascular Anesthesia, Meenakshi Multi Speciality Hospital, Tanjore, Tamil Nadu, India 2 Division of Cardiothoracic and Vascular Anesthesia, Sree Chitra Tirunal Institute for Medical Sciences and Technology, Trivandrum, Kerala, India
Click here for correspondence address and
email
Date of Submission | 21-Sep-2020 |
Date of Decision | 28-Apr-2021 |
Date of Acceptance | 16-May-2021 |
Date of Web Publication | 11-Apr-2022 |
|
|
 |
|
How to cite this article: Chennakeshavallu G N, Sruthi S, Babu S. Artifact in central venous pressure waveform due to central venous catheter tip abutting the wall of superior vena cava. Ann Card Anaesth 2022;25:243-4 |
How to cite this URL: Chennakeshavallu G N, Sruthi S, Babu S. Artifact in central venous pressure waveform due to central venous catheter tip abutting the wall of superior vena cava. Ann Card Anaesth [serial online] 2022 [cited 2022 May 27];25:243-4. Available from: https://www.annals.in/text.asp?2022/25/2/243/342844 |
To the Editor,
Malposition of central venous catheter (CVC) tip leads to an inconsistent waveform with erroneous central venous pressure (CVP) readings. We describe a case where the transduction of a malpositioned CVC tip displayed a waveform simulating an electrocardiographic rhythm.
A 60-year-old male with coronary artery disease was posted for minimally invasive direct coronary artery bypass grafting through left thoracotomy. After induction of anesthesia, trachea was intubated with a 37 Fr left-sided double lumen tube for lung isolation. Under ultrasound guidance, a triple-lumen CVC (7 Fr, 16 cm, Vygon, Germany) was placed in the left internal jugular vein (IJV) and fixed at 14 cm at skin level. On transducing the distal port, a waveform resembling electrocardiographic signal with pressure of 7 mm Hg was displayed [Figure 1]a. The correct position of CVC was confirmed with free aspiration of dark blood from all the three ports. Transesophageal echocardiography examination also revealed the presence of CVC in superior vena cava (SVC) [Figure 1]b. After completion of surgical procedure, patient was shifted to intensive care unit for postoperative management. The postoperative chest radiograph showed the tip of CVC abutting the lateral wall of SVC [Figure 1]c. On withdrawing the CVC by 1 cm, a normal CVP waveform was obtained [Figure 1d]. The postoperative course of the patient was uneventful. | Figure 1: (a) Multipara monitor showing central venous pressure waveform mimicking as electrocardiographic rhythm with p and QRS waves. (b) Transesophageal echocardiography imaging in midesophageal high bicaval view shows the presence of the CVC in the SVC. (c) Post-operative chest radiograph showing the CVC tip abutting against the lateral wall of SVC. (d) Multipara monitor showing return of normal central venous pressure waveform (a, v, x and y waves) after withdrawing the CVC by 1 cm. Abbreviations:- CVC- central venous catheter; ME, mid esophageal; RA, right atrium; RPA, right pulmonary artery; SVC – superior vena cava; TEE, transesophageal echocardiography
Click here to view |
Malposition of CVC tip is a relatively common complication.[1] Malposition is defined as placement of CVC in a vein other than the SVC, or impingement of CVC with the lateral wall of the SVC.[2] Placement of CVC under ultrasound guidance can even result in malposition of the CVC tip. Aspiration of venous blood and a good waveform with respiratory variations confirms the correct placement of CVC. The placement of CVC in left IJV has an increased chance of the catheter tip abutting the SVC wall as the brachiocephalic vein inserts at an angle into the SVC.[3] In our patient, the CVP waveform had a resemblance of electrocardiographic rhythm [Figure 1]a. The p wave (corresponding to actual “a” wave) in CVP waveform is because of increased transmural pressure in SVC during atrial contraction which led to CVC tip yielding away from the vessel wall. The artifactual QRS complex is because of fall in transmural pressure in SVC during atrial relaxation causing the CVC tip to abutt the vessel wall producing sharp spike in CVP waveform. Late perforation of the SVC after insertion of CVC through the left IJV has been reported.[4] Identification of the CVC tip grazing the wall of SVC is necessary to prevent SVC perforation. Analysis of artifactual CVP waveform and careful interpretation of chest radiograph may serve to identify the catheter tip abutting the wall of SVC, thereby avoiding untoward complications resulting from SVC perforation in the postoperative period.
Declaration of patient consent
The authors certify that they have obtained all appropriate patient consent forms. In the form the patient(s) has/have given his/her/their consent for his/her/their images and other clinical information to be reported in the journal. The patients understand that their names and initials will not be published and due efforts will be made to conceal their identity, but anonymity cannot be guaranteed.
Financial support and sponsorship
Nil.
Conflicts of interest
There are no conflicts of interest.
References | |  |
1. | Gibson F, Bodenham A. Misplaced central venous catheters, Applied anatomy and practical management. Br J Anaesth 2013;110(3):333-46. |
2. | Roldan CJ, Paniagua L. Central venous catheter intravascular malpositioning: Causes, prevention, diagnosis, and correction. West J Emerg Med 2015;1:658-64. |
3. | Mukau L, Talamini MA, Sitzmann JV. Risk factors for central venous catheter-related vascular erosions. JPEN J Parenter Enteral Nutr 1991;15(5):513-6. |
4. | Kurabe M, Watanabe T, Kohno T. Perforation of the superior vena cava 5 days after insertion of a central venous catheter through the left internal jugular vein. J Clin Anesth 2016;31:193-6. |

Correspondence Address: Saravana Babu Division of Cardiothoracic and Vascular Anesthesia, Sree Chitra Tirunal Institute for Medical Sciences and Technology, Trivandrum - 695 011, Kerala India
 Source of Support: None, Conflict of Interest: None  | Check |
DOI: 10.4103/aca.aca_241_20

[Figure 1] |
|
|
|
 |
 |
|
|
|