Comparative study of cardiac output measurement by regional impedance cardiography and thermodilution method in patients undergoing off pump coronary artery bypass graft surgery


Department of Cardiac Anaesthesiology, Medanta the Medicity, Gurugram, Haryana, India

Correspondence Address:
Amrita Guha
Milan Cooperative Society, Flat no. 801, Sec 39, Gurugram, Haryana . 122001
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/aca.aca_44_21

Rights and Permissions

Year : 2022  |  Volume : 25  |  Issue : 3  |  Page : 335-342

 

SEARCH
Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles

  Article in PDF (1,221 KB)
Email article
Print Article
Add to My List
Background: An ideal CO monitor should be noninvasive, cost effective, reproducible, reliable during various physiological states. Limited literature is available regarding the noninvasive CO monitoring in open chest surgeries. Aim: The aim of this study was to compare the CO measurement by Regional Impedance Cardiography (RIC) and Thermodilution (TD) method in patients undergoing off pump coronary artery bypass graft surgery (OPCAB). Settings and Design: We conducted a prospective observational comparative study of CO measurement by the noninvasive RIC method using the NICaS Hemodynamic Navigator system and the gold standard TD method using pulmonary artery catheter in patients undergoing OPCAB. A total of 150 data pair from the two CO monitoring techniques were taken from 15 patients between 40-70 years at various predefined time intervals of the surgery. Patients and Methods: We have tried to find out the accuracy, precision and cost effectiveness of the newer RIC technique. Mean CO, bias and precision were compared for each pair i.e.TD-CO and RIC-CO as recommended by Bland and Altman. The Sensitivity and specificity of cutoff value to predict change in TD-CO was used to create a Receiver operating characteristic or ROC curve. Results: Mean TD-CO values were around 4.52 ± 1.09 L/min, while mean RIC- CO values were around 4.77± 1.84 L/min. The difference in CO change was found to be statistically not significant (p value 0.667). The bias was small (-0.25). The Bland Altman plot revealed a mean difference of -0.25 litres. The RIC method had a sensitivity of 55.56 % and specificity of 33.33 % in predicting 15% change in CO of TD method and the total diagnostic accuracy was 46.67%. Conclusion: A fair correlation was found between the two techniques. The RIC method may be considered as a promising noninvasive, potentially low cost alternative to the TD technique of hemodynamic measurement.






[FULL TEXT] [PDF]*
 

 

 

 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 
 
 Reader Comments
 Email Alert *
  *
 * Requires registration (Free)
 
 ORIGINAL ARTICLE
 




Department of Cardiac Anaesthesiology, Medanta the Medicity, Gurugram, Haryana, India

Correspondence Address:
Amrita Guha
Milan Cooperative Society, Flat no. 801, Sec 39, Gurugram, Haryana . 122001
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/aca.aca_44_21

Rights and Permissions

Background: An ideal CO monitor should be noninvasive, cost effective, reproducible, reliable during various physiological states. Limited literature is available regarding the noninvasive CO monitoring in open chest surgeries. Aim: The aim of this study was to compare the CO measurement by Regional Impedance Cardiography (RIC) and Thermodilution (TD) method in patients undergoing off pump coronary artery bypass graft surgery (OPCAB). Settings and Design: We conducted a prospective observational comparative study of CO measurement by the noninvasive RIC method using the NICaS Hemodynamic Navigator system and the gold standard TD method using pulmonary artery catheter in patients undergoing OPCAB. A total of 150 data pair from the two CO monitoring techniques were taken from 15 patients between 40-70 years at various predefined time intervals of the surgery. Patients and Methods: We have tried to find out the accuracy, precision and cost effectiveness of the newer RIC technique. Mean CO, bias and precision were compared for each pair i.e.TD-CO and RIC-CO as recommended by Bland and Altman. The Sensitivity and specificity of cutoff value to predict change in TD-CO was used to create a Receiver operating characteristic or ROC curve. Results: Mean TD-CO values were around 4.52 ± 1.09 L/min, while mean RIC- CO values were around 4.77± 1.84 L/min. The difference in CO change was found to be statistically not significant (p value 0.667). The bias was small (-0.25). The Bland Altman plot revealed a mean difference of -0.25 litres. The RIC method had a sensitivity of 55.56 % and specificity of 33.33 % in predicting 15% change in CO of TD method and the total diagnostic accuracy was 46.67%. Conclusion: A fair correlation was found between the two techniques. The RIC method may be considered as a promising noninvasive, potentially low cost alternative to the TD technique of hemodynamic measurement.